External quality assessment for yaws elimination in low- and middle-income countries using plasmid-based proficiency test items
Por:
Mueller, C, Tchatchouang, S, Basing, LA, Ngazoa-Kakou, S, Hugues, KA, Handley, BL, Beiras, CG, Amanor, I, Ndzomo, P, Bakheit, MA, Becherer, L, Tabah, EN, Crucitti, T, Borst, N, Ries, C, Lueert, S, Frischmann, S, Gmoser, H, Landmann, E, Sylla, A, Kouamé-Sina, SM, Arhinful, D, Awondo, P, Burl, S, Harding-Esch, EM, Tano, A, Mitjà, O, Eyangoh, S, Addo, KK, Marks, M and Knauf, S
Publicada:
13 mar 2026
Resumen:
Background We aimed to establish an external quality assessment (EQA) programme for the yaws eradication campaign that would meet the needs of reference and district-level laboratories in low- and middle-income countries.Methodology/Principal Findings We designed proficiency testing items (PTIs) using a plasmid containing gene target sequences for Treponema pallidum (TP) and Haemophilus ducreyi (HD). The storage stability of the plasmids under different environmental conditions was then tested. A proficiency testing panel of seven swabs loaded with different concentrations of plasmids in different combinations, as well as human HEK293 cells to simulate the sample background, was prepared and sent to participating reference (RL) and district (DL) laboratories in Ghana, C & ocirc;te d'Ivoire and Cameroon followed by three rounds of blinded proficiency testing. We tested quantitative real-time PCR (qPCR) performance of reference laboratories and loop-mediated isothermal amplification (LAMP) performance of district laboratories and retested 20% of human field samples at the PTI provider's laboratory to further assess qPCR quality.Findings PTIs proved to be stable in dry conditions with no significant loss of copy number. Participating laboratories achieved qPCR results with a concordance of 95.0-100.0% (97.7% +/- 5.2% (mean +/- standard deviation ((SD)) with the provider and a concordance of 76.0-100.0% (TP: 90.3 +/- 13.7% and HD: 78.5 +/- 7.5% (mean +/- SD)) for LAMP results, with inconsistencies, particularly in the detection of low HD plasmid DNA levels combined with high TP plasmid copies. Retesting of field samples resulted in 100% correct TP and HD sample identification by the African reference laboratories.Conclusions/Significance We have developed a functional plasmid-based EQA programme specifically designed to meet the needs of resource-poor settings in the tropics. The programme is suitable as a blueprint for other disease programmes.
Filiaciones:
Mueller, C:
Fed Res Inst Anim Hlth, Inst Int Anim Hlth One Hlth, Friedrich Loeffler Inst, Greifswald, Insel Riems, Germany
Tchatchouang, S:
Ctr Pasteur Cameroun, NTD Platform, Yaounde, Cameroon
Basing, LA:
Univ Ghana, Noguchi Mem Inst Med Res, Accra, Ghana
Ngazoa-Kakou, S:
Inst Pasteur Cote Ivoire, Abidjan, Cote Ivoire
Hugues, KA:
Inst Pasteur Cote Ivoire, Abidjan, Cote Ivoire
Natl Program African Trypanosomiasis Eliminat, Abidjan, Cote Ivoire
Handley, BL:
London Sch Hyg & Trop Med, Clin Res Dept, London, England
:
Hosp Badalona Germans Trias & Pujol, Flight Infect Dis Fdn, Skin, Barcelona, Spain
Univ Autonoma Barcelona, Bellaterra, Cerdanyola Del, Spain
Amanor, I:
Univ Ghana, Noguchi Mem Inst Med Res, Accra, Ghana
Ndzomo, P:
Ctr Pasteur Cameroun, NTD Platform, Yaounde, Cameroon
Bakheit, MA:
Mast Diagnost GmbH, Reinfeld, Germany
Becherer, L:
Univ Freiburg, IMTEK Dept Microsyst Engn, Lab MEMS Applicat, Freiburg, Germany
Tabah, EN:
Minist Publ Hlth, Natl Buruli Ulcer Leprosy Yaws & Leishmaniasis Con, Yaounde, Cameroon
Univ Dschang, Fac Med & Pharmaceut Sci, Dept Publ Hlth, Dschang, Cameroon
Crucitti, T:
Inst Pasteur Madagascar, Expt Bacteriol Unit, Antananarivo, Madagascar
Borst, N:
Univ Freiburg, IMTEK Dept Microsyst Engn, Lab MEMS Applicat, Freiburg, Germany
Hahn Schickard, Freiburg, Germany
Ries, C:
Fed Res Inst Anim Hlth, Inst Int Anim Hlth One Hlth, Friedrich Loeffler Inst, Greifswald, Insel Riems, Germany
Lueert, S:
Fed Res Inst Anim Hlth, Inst Int Anim Hlth One Hlth, Friedrich Loeffler Inst, Greifswald, Insel Riems, Germany
Frischmann, S:
Mast Diagnost GmbH, Reinfeld, Germany
Gmoser, H:
Univ Freiburg, IMTEK Dept Microsyst Engn, Lab MEMS Applicat, Freiburg, Germany
Landmann, E:
Mast Diagnost GmbH, Reinfeld, Germany
Sylla, A:
Inst Pasteur Cote Ivoire, Abidjan, Cote Ivoire
Kouamé-Sina, SM:
Inst Pasteur Cote Ivoire, Abidjan, Cote Ivoire
Arhinful, D:
Univ Ghana, Noguchi Mem Inst Med Res, Accra, Ghana
Awondo, P:
Ctr Pasteur Cameroun, NTD Platform, Yaounde, Cameroon
Burl, S:
London Sch Hyg & Trop Med, Clin Res Dept, London, England
Harding-Esch, EM:
London Sch Hyg & Trop Med, Clin Res Dept, London, England
Tano, A:
Inst Pasteur Cote Ivoire, Abidjan, Cote Ivoire
:
Hosp Badalona Germans Trias & Pujol, Flight Infect Dis Fdn, Skin, Barcelona, Spain
Univ Autonoma Barcelona, Bellaterra, Cerdanyola Del, Spain
Eyangoh, S:
Ctr Pasteur Cameroun, NTD Platform, Yaounde, Cameroon
Addo, KK:
Univ Ghana, Noguchi Mem Inst Med Res, Accra, Ghana
Marks, M:
London Sch Hyg & Trop Med, Clin Res Dept, London, England
Hosp Trop Dis, London, England
Knauf, S:
Fed Res Inst Anim Hlth, Inst Int Anim Hlth One Hlth, Friedrich Loeffler Inst, Greifswald, Insel Riems, Germany
Justus Liebig Univ Giessen, Fac Vet Med, Giessen, Germany
Green Submitted, gold
|